Abstract
We present the first example of a microfabricated PCR thermocycling chip integrated with advanced sample pre-treatment. The device consists of a SU-8 based PCR thermocycler chip monolithically integrated with dielectrophoretic (DEP) sample pre-treatment. This allows on chip purification of samples by removing inhibitors that would otherwise prevent PCR amplification. Integrated thin film heaters and a resistive temperature sensor control the PCR thermocycling. The sample pre-treatment system uses a combination of fluid flow and cell capturing by DEP to perform the sample treatment. The sample pre-treatment system was tested on a sample containing yeast cells and heparin that inhibits PCR. Only treated sample could be successfully amplified using PCR.
| Originalsprog | Engelsk |
|---|---|
| Titel | TRANSDUCERS 2003 - 12th International Conference on Solid-State Sensors, Actuators and Microsystems, Digest of Technical Papers |
| Antal sider | 4 |
| Udgivelsessted | Boston, USA |
| Publikationsdato | 2003 |
| Sider | 214-217 |
| Artikelnummer | 1215291 |
| ISBN (Elektronisk) | 0780377311, 9780780377318 |
| DOI | |
| Status | Udgivet - 2003 |
| Udgivet eksternt | Ja |
| Begivenhed | 12th International Conference on Solid-State Sensors, Actuators and Microsystems, TRANSDUCERS 2003 - Digest of Technical Papers - Boston, USA Varighed: 8 jun. 2003 → 12 jun. 2003 Konferencens nummer: 12 |
Konference
| Konference | 12th International Conference on Solid-State Sensors, Actuators and Microsystems, TRANSDUCERS 2003 - Digest of Technical Papers |
|---|---|
| Nummer | 12 |
| Land/Område | USA |
| By | Boston |
| Periode | 08/06/03 → 12/06/03 |
Fingeraftryk
Dyk ned i forskningsemnerne om 'Microfabricated DNA amplification device monolithically integrated with advanced sample pre-treatment'. Sammen danner de et unikt fingeraftryk.Citationsformater
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver