Accuracy of radiographic measurements of fracture-induced deformity in the distal radius

  • Janni Jensen
  • , Ole Graumann
  • , Oke Gerke
  • , Trine Torfing
  • , Helle Precht
  • , Benjamin Schnack Rasmussen
  • , Hans B. Tromborg
  • University of Southern Denmark
  • Aarhus Univeristy Hospital
  • Department of Clinical Immunology and Biochemistry, Lillebaelt Hospital, Vejle, Denmark.

Research output: Contribution to journalJournal articleResearchpeer-review

Abstract

BackgroundManagement of the distal radius fracture (DRF) is to some extent based on radiographic characterization of fracture displacement. It remains unclear, however, if the measurements used to quantify displacement are accurate.
PurposeTo quantify accuracy of two radiographic measurements: dorsal/volar tilt and fracture compression, measured indirectly as ulnar variance (UV), using radiostereometric analyses (RSA) as reference standard.
Material and MethodsTwenty-one fresh frozen non-fractured human cadaveric forearms (right = 11, left = 10) were thawed and eligible for inclusion. The forearms were mounted on a custom made platform that allowed for controlled forearm rotation, and they underwent two rounds of imaging (both rounds consisted of RSA and radiographs). In round one, the non-fractured forearms were radiographed. In round two, artificial DRF´s with compression and dorsal angulation were created and imaging procedures repeated. Change in tilt and UV between the non-fractured and later fractured forearms was defined as fracture-induced deformity. Deformity was measured radiographically and additionally calculated using RSA. Bland Altman analyses were used to estimate agreement between radiographically measured, and RSA calculated, fracture-induced deformity.
ResultsOur results indicated that radiographs underestimate the amount of fracture-induced deformity. Mean measured differences (bias) in dorsal tilt deformity between radiographs and RSA were −2.5° for both observers. The corresponding values for UV were −1.4 mm and −1.5 mm.
ConclusionQuantifying fracture-induced deformity on radiographs underestimated the actual deformity when compared to RSA calculated deformity. These findings suggest that clinicians, at least in part, base fracture management and potentially corrective surgery on inaccurate measurements.