Skip to main navigation Skip to search Skip to main content

Comparison of conventional hand examination on six optimised DR systems

  • University of Southern Denmark
  • Sydvestjysk Sygehus
  • Norwegian University of Science and Technology

Research output: Contribution to journalJournal articleResearchpeer-review

Abstract

The purpose of this study was to investigate the challenges in comparing digital radiography (DR) systems from different vendors for various combinations of exposure factors in posterior–anterior hand radiographs. Image quality was evaluated for a range of tube voltages and tube current-time products using a technical contrast-detail (CDRAD) phantom and an anthropomorphic hand phantom. 900 technical CDRAD images were analysed providing quality figures of merit (IQFinv) and two experienced reporting radiographers using visual grading analysis (VGA) scored 108 anthropomorphic images. This study demonstrates the differences between the DR systems included. When compensating for variations in dose, Canon showed superior results for technical image quality and Fuji for visual image quality for a standard dose point at DR hand examination (ln(DAP) 1.1, 50 kV and 2.5 mAs).
Original languageEnglish
JournalRadiation Protection Dosimetry
Pages (from-to)27-35
ISSN0144-8420
DOIs
Publication statusPublished - 2021

Keywords

  • radiography

Fingerprint

Dive into the research topics of 'Comparison of conventional hand examination on six optimised DR systems'. Together they form a unique fingerprint.