Intra- and interobserver reliability of quantitative ultrasound measurement of the plantar fascia

  • Michael Skovdal Rathleff
  • , Carsten Møller Mølgaard
  • , Jens Lykkegaard Olesen
  • Aalborg Universitetshospital

Research output: Contribution to journalJournal articleResearchpeer-review

Abstract

Purpose: To determine intra- and interobserver reliability and measurement precision of sonographic assessment of plantar fascia thickness when using one, the mean of two, or the mean of three measurements. Methods: Two experienced observers scanned 20 healthy subjects twice with 60 minutes between test and retest. A GE LOGIQe ultrasound scanner was used in the study. The built-in software in the scanner was used to measure the thickness of the plantar fascia (PF). Reliability was calculated using intraclass correlation coefficient (ICC) and limits of agreement (LOA). Results: Intraobserver reliability (ICC) using one measurement was 0.50 for one observer and 0.52 for the other, and using the mean of three measurements intraobserver reliability increased up to 0.77 and 0.67, respectively. Interobserver reliability (ICC) when using one measurement was 0.62 and increased to 0.82 when using the average of three measurements. LOA showed that when using the average of three measurements, LOA decreased to 0.6 mm, corresponding to 17.5% of the mean thickness of the PF. Conclusions: The results showed that reliability increases when using the mean of three measurements compared with one. Limits of agreement based on intratester reliability shows that changes in thickness that are larger than 0.6 mm can be considered actual changes in thickness and not a result of measurement error.

Original languageEnglish
JournalJournal of Clinical Ultrasound
Volume39
Issue number3
Pages (from-to)128-134
Number of pages7
ISSN0091-2751
DOIs
Publication statusPublished - 2011

Keywords

  • physiotherapy

Fingerprint

Dive into the research topics of 'Intra- and interobserver reliability of quantitative ultrasound measurement of the plantar fascia'. Together they form a unique fingerprint.

Cite this